Patient dose in image guided radiotherapy: Monte Carlo study of the CBCT dose contribution

Autori

  • Salvatore Leotta Radiosurgery Center, Agropoli
  • Ernesto Amato Dr. Ernesto Amato, PhD<br />Università degli Studi di Messina<br />Dipartimento di Scienze Biomediche e delle Immagini Morfologiche e Funzionali<br />Sezione di Scienze Radiologiche<br />tel: +39 090 2212942<br />email: eamato@unime.it http://orcid.org/0000-0001-7416-8824
  • Nicola Settineri Azienda Ospedaliera Papardo, Messina
  • Emilia Basile Azienda Ospedaliera Papardo, Messina
  • Antonio Italiano Istituto Nazionale di Fisica Nucleare, Sezione di Catania
  • Lucrezia Auditore Azienda Ospedaliera Universitaria “G. Martino”, Messina
  • Anna Santacaterina Azienda Ospedaliera Papardo, Messina
  • Stefano Pergolizzi Università degli Studi di Messina

DOI:

https://doi.org/10.1478/AAPP.961A3

Parole chiave:

Image Guided Radiotherapy, Monte Carlo, GAMOS, Geant4, organ dose

Abstract

Image Guided RadioTherapy (IGRT) is a technique whose diffusion is growing thanks to the well-recognized gain in accuracy of dose delivery. However, multiple Cone Beam Computed Tomography (CBCT) scans add dose to patients, and its contribution has to be assessed and minimized. Aim of our work was to evaluate, through Monte Carlo simulations, organ doses in IGRT due to CBCT and therapeutic MV irradiation in head-neck, thorax and pelvis districts. We developed a Monte Carlo simulation in GAMOS (Geant4-based Architecture for Medicine-Oriented Simulations), reproducing an Elekta Synergy medical linac operating at 6 and 10 MV photon energy, and we set up a scalable anthropomorphic model. After a validation by comparison with the experimental quality indexes, we evaluated the average doses to all organs and tissues belonging to the model for the three cases of irradiated district. Scattered radiation in therapy is larger than that diffused by CBCT by one to two orders of magnitude.

Biografie autore

  • Ernesto Amato, Dr. Ernesto Amato, PhD<br />Università degli Studi di Messina<br />Dipartimento di Scienze Biomediche e delle Immagini Morfologiche e Funzionali<br />Sezione di Scienze Radiologiche<br />tel: +39 090 2212942<br />email: eamato@unime.it
    Dipartimento di Scienze Biomediche, Odontoiatriche e delle Immagini Morfologiche e Funzionali
  • Stefano Pergolizzi, Università degli Studi di Messina
    Dipartimento di Scienze Biomediche, Odontoiatriche e delle Immagini Morfologiche e Funzionali

Pubblicato

2018-02-23

Fascicolo

Sezione

Articoli