On the need of ad hoc calibration factors for activity measurement of solutions containing radioactive microspheres
DOI:
https://doi.org/10.1478/AAPP.1022A6Parole chiave:
activity measurements, calibration factors, glass microspheres, resin microspheresAbstract
Microspheres labelled with radioactive nuclides are widely used for TransArterial RadioEmbolization (TARE) in the treatment of patients with hepatocellular carcinoma (HCC) and/or hepatic metastases. Nowadays, three commercially available devices can be used: 90Y-loaded glass (Theraspheres®, Boston Scientific, Marlborough, MA, USA) and resin microspheres (SirSpheres®, SIRTex medical, Woburn, MA, USA), or 166Ho-microspheres (QuiremSpheresTM, Quirem Medical, Deventer, The Netherlands). With reference to 90Y-loaded glass and resin microspheres, recent studies (S. A. Graves et al., J. Nucl. Med. 23, 1131-1135 (2022); S. Gnesin et al., J. Nucl. Med. 64, 825-828 (2023)) highlighted the difference between the measured activity of commercial vials and the nominal one declared by the vendors. As discussed by L. Auditore et al. (J. Nucl. Med. 64, 1471-1477 (2023)), this discrepancy can be attributed to the procedure used to measure the activity of the vials at the vendor sites and to neglect the emission of Internal Bremsstrahlung (IB) photons among the radioactive decay particles emitted by 90Y. The aim of this study was to investigate, by means of Monte Carlo simulations, the causes of such discrepancies, focusing on the geometrical factors affecting the activity measurement. In this study an activity calibrator was reproduced in Monte Carlo simulation environment and the measurement of the activity of a vial containing YCl3 or 90Y-labelled microspheres was simulated. With reference to the glass microspheres, the relative difference between the signal produced by the two solutions, YCl3 and 90Y-labelled microspheres, is 23.5%. This discrepancy reduces to −1.7% if the 90Y-labelled microspheres are simulated as uniformly distributed in the vial. Moreover, results indicate that the geometry of the vial affects the results reducing the discrepancy from 23.5% to 16.6%. Similar results were obtained for 90Y-resin microspheres for which the discrepancy with respect to the reference solution reduces from −15% to −8.6% when considering the microspheres homogeneously dispersed in the vial. Since the activity of a device employed for therapy has to be accurately known, this study highlights the need of updating the procedure of measuring the activity of such particulate solutions by using ad hoc calibration factors to consider the real vial geometry and distribution of the radioactive solute in the vial.Dowloads
Pubblicato
2024-11-12
Fascicolo
Sezione
Articoli
Licenza

This work is licensed under a Creative Commons Attribution 4.0 International License.
- Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).